Slit Lamp Illuminator | ||
Slit image width | 0 – 8 mm continuous | |
Slit image length | 1 -8 mm continuous | |
Illumination field diameter |
8, 5, 3, 2, 1, 0.2 mm | |
Slit image radial range | ± 90° | |
Radial movement of the slit light | horizontal ± 90°, vertical 0 - 20° | |
Filters | blue, redfree (green), grey (10 %), heat absorption filter | |
Light source | Tungsten incandescent light bulb 6 V / 4.5 A | |
Illumination intensity | 150 000 Lux at position 1/2 | |
Working temperature | 0° to 50° C | |
Stereo-Microscope | ||
Objective lens focal length | 105 mm | |
Magnification changer | 6.3x 10x 16x 25x 40x | |
Ocular magnification | 12.5x | |
Range of adjusting eye-pieces | + 8 to -8 diopters | |
Inter pupillary distance | 52 - 78 mm | |
Instrument base | ||
Operation | Single handed 3-dimensional operation of the guide lever | |
Adjustment of the instrument base | 80 mm (length), 30 mm (height), 100 mm (side) | |
Electric supply (class II, type B) | 85 to 264 VAC , 47 to 63 Hz, 110VA | |
Weight | 12.76 kg (without power supply, headrest and options) |